Organic Volatile Impurities (USP 467)
Method: | GC |
Matrix: | Standards |
Application No.: | 101832 |
Columns: | DM-624 30 m x 0.53 mm x 3.00 μm |
Cat No.: | 7751 |
Conditions: | Oven Temp.: 35 ºC (hold 10 min) to 100 ºC at 5 ºC/min to 240 ºC (hold 5 min) at 25 ºC/min Carrier Gas: He, 35 cm/sec, 35 ºC Injection: Split, 2:1, 220 ºC Detector: FID, 1.05 x 10-11 AFS, 240 ºC |
Author: | Dikma Technologies Inc. |
Keyword: | Organic Volatile Impurities, GC, DM-624 |
|
|
Table: | 1. Ethylene oxide 2. Methanol 3. Ethanol 4. Diethyl ether 5. 1,1-Dichloroethene 6. Acetone 7. Isopropanol 8. Acetonitrile 9. Methylene chloride 10. n-Hexane 11. n-Propanol 12. Methyl ethyl ketone 13. Ethyl acetate 14. Tetrahydrofuran 15. Chloroform 16. 1,1,1-Trichloroethane 17. Carbon tetrachloride 18. Benzene 19. 1,2-Dichloroethane 20. Heptane 21. Trichloroethylene 22. n-Butanol 23. 1,4-Dioxane 24. Pyridine 25. Toluene |