Determination of Migration of Bisphenol A (BPA) from Plastic Baby Bottles
Method: | SPE |
Matrix: | Plastic |
Application No.: | 101755 |
Columns: | ProElut PLS Glass Cartridge, 200 mg / 6 mL 30/pk |
Cat No.: | 68012G |
Sample Pretreatment: | 1. Sample purification Wash bottles, dry completely. Add distilled water so that each 8 m2 of plastic contact area corresponds to 10 mL of simulant immersion. Sealed with aluminum foil, place in oven at 100 ºC for 1 h. Cool to room temperature, then transfer to glass bottles, and seal until detection. 2. Sample purification ProElut™ PLS GLASS 200 mg / 6 mL (Cat#68012G) Condition: 6 mL MeOH / 6 mL H2O Load: Adding sample Wash: 6 mL 5 % MeOH in H2O Elute: 6 mL MeOH Reconstitute: Evaporate to near dry at 40 ºC, reconstitute to 1 mL with mobile phase |
Conditions: | HPLC method Column: Inspire™ 5 µm C18, 250 x 4.6 mm (Cat#81006) Mobile Phase: MeCN:2% CH3COOH in H2O = 40:60 Flow Rate: 1.0 mL/min Temperature: 30 ºC Detection: FLD Ex: 227 nm, Em: 313 nm Injection Volume: 20 µL |
Author: | Dikma Technologies Inc. |
Keyword: | Migration of Bisphenol A (BPA) from Plastic Baby Bottles, Bisphenol A, BPA, Inspire, SPE, PLS |
|
Contact Dikma:
USA
51 Massier Lane
Foothill Ranch, CA 92610, USA
Tel: 1-866-889-9072
Fax: 1-866-833-2653
Email: sale@dikmatech.com
Business hours: 9 AM - 5 PM PST
Canada
255 Shields Court, Unit A
Markham, ON L3R 8V2, Canada
Tel: 905-944-8066
Fax: 905-944-0181
Toll-Free:1-866-889-9072
Email: sales@dimaglass.com
Business hours: 9 AM - 5 PM EST
Asia / Pacific Area
Room 9, 5F., No.763 Wenlin Road,
Shilin District, Taipei City
111, Taiwan
Email: paulw@dikmatech.com