Determination of Tetracyclines in Serum
Method: | SPE |
Matrix: | Serum |
Application No.: | 101752 |
Columns: | ProElut PLS 60 mg / 3 mL 50/pk |
Cat No.: | 68003 |
Sample Pretreatment: | 1. Sample preparation 2 mL serum, add 40 µL H3PO4 2. Sample purification ProElut™ PLS 60 mg / 3 mL (Cat#68003) Condition: 3 mL MeOH / 3 mL H2O Load: 2 mL sample Wash: 3 mL 5 % MeOH in H2O Elute: 3 mL MeOH Reconstitute: Evaporate to near dry at 30 ºC, reconstitute to 1 mL with mobile phase |
Conditions: | HPLC method Column: Inspire™ 5 µm C18, 250 x 4.6 mm (Cat#81006) Mobile Phase: MeOH:MeCN:10 mM oxalic acid in H2O = 15:15:70 Flow Rate: 1.0 mL/min Temperature: 30 ºC Detection: UV 365 nm Injection Volume: 20 µL |
Author: | Dikma Technologies Inc. |
Keyword: | Tetracyclines, Serum, Inspire, SPE, PLS |
|
Contact Dikma:
USA
51 Massier Lane
Foothill Ranch, CA 92610, USA
Tel: 1-866-889-9072
Fax: 1-866-833-2653
Email: sale@dikmatech.com
Business hours: 9 AM - 5 PM PST