Determination of Phenols in Water
Method: | SPE |
Matrix: | Water |
Application No.: | 101748 |
Columns: | ProElut PLS 60 mg / 3 mL 50/pk |
Cat No.: | 68003 |
Sample Pretreatment: | 1. Sample preparation 100 mL sample, adjust to pH 2 with H3PO4 2. Sample purification ProElut™ PLS 60 mg / 3 mL (Cat#68003) Condition: 3 mL MeOH:MTBE = 10:90 / 3 mL MeOH / 3 mL H2O Load: 100 mL sample, flow rate ≤ 5 mL/min Wash: 3 mL H2O Elute: 3 mL MeOH:MTBE = 10:90 Reconstitute: Evaporate at 40 ºC by N2, reconstitute to 1 mL with MeCN:H2O (50:50, V / V) solution |
Conditions: | HPLC method Column: Inspire™ 5 µm C18, 150 x 4.6 mm (Cat#81001) Mobile Phase: A: 1% CH3COOH in H2O, B: 1% CH3COOH in MeCN Flow Rate: 1.0 mL/min Temperature: Ambient Detection: UV 280 nm Injection Volume: 20 µL |
Author: | Dikma Technologies Inc. |
Keyword: | Phenols, SPE, PLS, Inspire |
|