Glycols / Alcohols
Method: | GC |
Matrix: | Standards |
Application No.: | 101139 |
Columns: | DM-5 30 m x 0.53 mm x 1.00 μm |
Cat No.: | 7249 |
Conditions: | Oven Temp.: 40 ºC to 185 ºC (hold 5 min) at 15 ºC/min Carrier Gas: He, 40 cm/sec Injection: Direct, 150 ºC Sample: Glycols and alcohols, 1.0 μL, 100 ppm Detector: FID, 8 x 10-11 AFS, 200 ºC |
Author: | Dikma Technologies Inc. |
Publisher: | CCR00327 |
Keyword: | Glycols, Alcohols, GC, DM-5, petrochemical |
|
|
Table: | 1. n-Butyl alcohol 2. Ethylene glycol 3. Propylene glycol 4. Butyl cellosolve 5. Butyl carbitol |